The Acquity system comes complete with OQ Certification.
Note: Our warranty terms quoted for this system apply to mainland UK customers. For clients outside of this locale, please enquire at point of sale.
hplc4sale.co.uk (Jaytee Biosciences Ltd) warranty all reconditioned instruments for a pre-agreed period. This period can be from 30 days to 1 year from the date of installation / delivery of the instrument. Generally newer equipment has a longer warranty than older equipment.
Additional warranty periods can be negotiated at the time of purchase if required.
The list below outlines the items covered by our warranty:
HPLC Pumps
HPLC Detectors
Autosampler
GC Autosampler
Gas Chromatograph
Misceallenous
Acquity Binary UPLC system (H-Class) from Waters. This fantastic system delivers the flexibility of quaternary solvent blending with the advanced performance of UPLC® separations. The systems holistic design is targeted for routine analysis and method development use and is perfectly suited for running both HPLC and UPLC applications while still delivering the improved resolution and sensitivity of UPLC separations. Complete with an OQ certification - see photos.
Includes fully reconditioned and serviced:
Optional extras (enquire for price):
We can supply software, PC printer and installation - please call +44(0)1227 265333 for a full quotation. Contact Us
Dwell volume (total system): <400 µL (includes standard 100 µL mixer)
Integrated leak management: Leak sensors, as standard, and safe leak handling.
Quantum synchronization: Injection synchronization between pump and sample manager enhances retention time reproducibility.
pH range: 2 to 12
Unattended operation: Leak sensors and safe leak handling, full 96-hour diagnostic data display through console software.
Cycle time: <30 s inject-to-inject
Settable flow rate range: 0.010 to 2.000 mL/min, in 0.001 mL increments (firmware version 1.5x and earlier).
Maximum operating pressure: 15,000 psi up to 1.0 mL/min,7800 psi up to 2.2 mL/min (firmware version 1.6x and later).
Number of solvents: Blend up to four solvents in any combination (standard). Expanded solvent choices with optional six-port solvent select valve.
Solvent degassing: Integrated vacuum degassing, four chambers. One additional chamber for the SM-FTN purge solvent.
Solvent blending: Automated, on-line pH, ionic strength, and organic modifier blending from pure solvents with Auto•Blend Plus™ Technology.
Gradient formation: Low-pressure mixing, quaternary gradient.
Flow accuracy: ±1.0% at 0.5 to 2.0 mL/min using 100% A (with i2 Valve). Back pressure 600 to 1000 psi with degassed H20.
Injection volume range: 0.1 to 10.0 µL as standard in 0.1 µL increments. Up to 1000.0 µL with optional extension loops.
Accuracy (aspiration): ±0.2 µL, measured by fluid weight removed from vial with 10.0 µL injections averaged over 20 injections using standard 100 µL syringe.
Linearity: >0.999 (standard needle) caffeine 0.030 mg/mL, ACN/H2O (10:90), isocratic 0.6 mL/min, 0.2 to 10.0 µL, 1 to 70% needle volume.
Sample compartment: 4.0 to 40.0 °C, settable in 0.1 °C increments
Wavelength range 1: 190 to 800 nm
Wavelength accuracy: ±1 nm
Linearity range: Deviation at 2.0 AU ≤ 5%, propylparaben, at 257 nm
Data acquisition: Up to 80 Hz.
Pressure limit: 1000 psi
Flow cell design: Light-guiding UPLC intelligent technology flow cell.
Column capacity: Single column, up to 4.6 mm internal diameter (I.D.), up to 300 mm lengthwith filter or guard column.
Column compartment temperature range: 20.0 to 90.0 °C, settable in 0.1 °C increments.
Column compartment temperature accuracy: ±0.5 °C at sensor.
Column compartment temperature stability: ±0.3 °C at sensor.
Solvent conditioning: Active pre-heating as standard.
Column tracking: eCord™ Technology column information management tracks and archives column usage history.